A client with an implantable cardioverter-defibrillator develops ventricular tachycardia

Journal Article

Joshua M. Cooper,

Cardiovascular Division, Department of Internal Medicine

,

University of Pennsylvania Health System

,

3400 Spruce Street, 9 Founders Pavilion, Philadelphia, PA 19104

,

USA

Search for other works by this author on:

Nimrod Lavi,

Cardiovascular Division, Department of Internal Medicine

,

University of Pennsylvania Health System

,

3400 Spruce Street, 9 Founders Pavilion, Philadelphia, PA 19104

,

USA

Search for other works by this author on:

Glenn M. Polin,

Cardiovascular Division, Department of Internal Medicine

,

University of Pennsylvania Health System

,

3400 Spruce Street, 9 Founders Pavilion, Philadelphia, PA 19104

,

USA

Search for other works by this author on:

Sanjay Dixit,

Cardiovascular Division, Department of Internal Medicine

,

University of Pennsylvania Health System

,

3400 Spruce Street, 9 Founders Pavilion, Philadelphia, PA 19104

,

USA

Search for other works by this author on:

Ralph J. Verdino

Cardiovascular Division, Department of Internal Medicine

,

University of Pennsylvania Health System

,

3400 Spruce Street, 9 Founders Pavilion, Philadelphia, PA 19104

,

USA

Search for other works by this author on:

Published:

24 November 2010

  • PDF
  • Split View
    • Article contents
    • Figures & tables
    • Video
    • Audio
    • Supplementary Data
  • Cite

    Cite

    Joshua M. Cooper, Nimrod Lavi, Glenn M. Polin, Sanjay Dixit, Ralph J. Verdino, Temporary external implantable cardioverter defibrillator in the pacemaker-dependent ventricular tachycardia patient, EP Europace, Volume 13, Issue 5, May 2011, Pages 761–763, //doi.org/10.1093/europace/euq412

    Close

    • Email
    • Twitter
    • Facebook
    • More

Close

Navbar Search Filter Microsite Search Term Search

Abstract

A patient with ischaemic cardiomyopathy underwent implantable cardioverter defibrillator (ICD) extraction for a severe pocket infection and sepsis. During 5 weeks of critical medical care after device extraction, heart block and recurrent monomorphic ventricular tachycardia (VT) were managed with an ‘externalized' active fixation pacemaker lead and a resterilized ICD generator. This case demonstrates how a permanent pacing lead and an external ICD generator can provide reliable temporary pacing and automatic anti-tachycardia pacing for recurrent VT until a new device can be implanted and more permanent VT treatment options are feasible.

Case

When an implanted pacemaker or implantable cardioverter defibrillator (ICD) becomes infected, the entire system, including the pulse generator and all leads, must be removed for infection resolution. The timing of permanent device re-implantation is dependent upon infection severity, the effectiveness of antimicrobial treatment, and the documentation of sterile blood cultures. In a pacemaker-dependent patient, temporary pacing is needed until the new device system can be implanted. An active fixation permanent pacemaker lead can be percutaneously implanted via the internal jugular or subclavian vein and attached to a resterilized, external permanent pacemaker for temporary right ventricular (RV) pacing.1 This technique is associated with fewer adverse events than other temporary pacing modalities and is cost-effective.2,3 In ICD patients with recurrent ventricular tachycardia (VT) after device extraction, tachyarrhythmias also need to be managed.

A 74-year-old man with ischaemic cardiomyopathy was transferred to our institution with a purulent pocket infection and severe sepsis with Staphylococcus aureus, 3 weeks after an ICD generator change. His dual-chamber ICD was implanted 9 years prior for VT, and he developed complete heart block in the interim, leaving him pacemaker-dependent. His left ventricular ejection fraction was 30%, and, since the recent ICD replacement, he had 81 episodes of asymptomatic monomorphic VT (cycle length 370–430 ms), despite chronic treatment with amiodarone. Each episode was successfully terminated with a single anti-tachycardia pacing (ATP) burst, and no shocks were delivered.

On presentation, he was ill-appearing, with persistent methicillin-sensitive S. aureus bacteraemia and fevers to 103°F, despite treatment with intravenous nafcillin. The patient underwent device removal, extensive debridement of the grossly purulent left pectoral device pocket, and laser extraction of his chronic atrial and ventricular leads. A femoral quadripolar catheter was used for temporary pacing during the procedure. The right internal jugular vein site was then sterilely prepped, an active fixation permanent pacemaker lead was percutaneously implanted in the RV apex, and the lead was connected to an external resterilized ICD generator (Figure 1). The ventricular pacing threshold was 0.4 V at 0.5 ms. The device was programmed for bradycardia pacing at VVI 80 p.p.m., and a VT zone of 140–250 b.p.m. was set, with six burst ATP therapies only.

Figure 1

(Top, left) Chest radiograph of temporary pacing system. An active fixation permanent pacemaker lead is affixed to the RV apex (black arrow) and is connected to an external ICD (black asterisk), which is secured to the neck with an adhesive dressing. (Top, right) Photograph of external temporary pacing system at the time of removal. The pacemaker lead is sutured to the skin via its suture sleeve at the right internal jugular vein access site (white arrow) and is connected to a resterilized ICD (white asterisk). (Bottom) Sample telemetry strip from a VT event after extraction of the infected ICD system. Ventricular tachycardia is successfully detected and treated by the temporary pacing/external ICD system with a single 10-pulse burst of ATP. VVI pacing at 80 p.p.m. is seen after VT termination, with heart block and dissociated P-waves noted.

The patient remained intubated after the extraction procedure because of S. aureus pneumonia, and a tracheostomy was necessary after 2 weeks because of ongoing dependence on mechanical ventilation. He required ongoing pressor support due to sepsis-induced hypotension in the setting of severe cardiomyopathy, and his critical clinical status precluded a VT ablation procedure. Over a 5-week period, the temporary device system provided reliable pacing and multiple successful ATP therapies for recurrent monomorphic VT, and no external shocks were needed. The patient recovered from his infection, and a new permanent right-sided biventricular ICD was implanted in the right pectoral region.

Conflict of interest: J.M.C., S.D., and R.J.V. have modest conflicts of interest to report with Medtronic, Boston Scientific, St Jude, and Biotronik in the form of modest honoraria for educational programmes and training support for the institution's fellowship programme.

References

1

,  ,  . 

Modified temporary cardiac pacing using transvenous active fixation leads and external re-sterilized pulse generators

,

J Am Coll Cardiol

,

2006

, vol.

47

 

(pg.

1487

-

9

)

2

,  ,  ,  ,  ,  , et al. 

Percutaneous lead implantation connected to an external device in stimulation-dependent patients with systemic infection—a prospective and controlled study

,

Pacing Clin Electrophysiol

,

2006

, vol.

29

 

(pg.

875

-

9

)

3

,  ,  ,  ,  ,  , et al. 

Utility and cost effectiveness of temporary pacing using active fixation leads and an externally placed reusable permanent pacemaker

,

Am J Cardiol

,

2006

, vol.

98

 

(pg.

1613

-

5

)

Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2010. For permissions please email:

Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2010. For permissions please email:

Topic:

  • artificial cardiac pacemaker
  • tachycardia
  • sepsis
  • tachycardia, ventricular
  • heart block
  • ventricular tachycardia, monomorphic
  • pacemaker leads
  • temporary cardiac pacemaker procedure
  • implantable defibrillators
  • ischemic cardiomyopathy
  • infections
  • medical devices
  • generators

Citations

Views

Altmetric

Email alerts

More on this topic

Related articles in PubMed

Citing articles via

  • Latest

  • Most Read

  • Most Cited

More from Oxford Academic

What is the most common complication of defibrillator?

The main risks related to ICD implantation surgery include:.
bleeding..
infection..
pneumothorax..
damage to the heart or to a blood vessel..

What are nursing interventions for ventricular tachycardia?

Patients with pulseless ventricular tachycardia are treated the same as those with ventricular fibrillation, meaning they require immediate cardiopulmonary resuscitation (CPR) and defibrillation. Once the rhythm converts, the patient will need an infusion of amiodarone or lidocaine.

What is the appropriate treatment for ventricular tachycardia?

If you have ventricular tachycardia, you may be given medications called anti-arrhythmics by mouth or IV to slow the fast heart rate. Other heart medications, such as calcium channel blockers and beta blockers, may be prescribed with anti-arrhythmic drugs.

Can a pacemaker cause ventricular tachycardia?

Pacemaker-induced ventricular tachycardia is an uncommon complication that can happen any time after the implantation of the device. Period may range from a few days to years after the device implantation. Runaway pacemaker, if happens due to malfunction of the device, results in ventricular tachycardia.

Chủ đề